Technical data
Ventilation modes: | IPPV; A/C; SIMV; |
Ventilator parameter ranges | |
Tidal volume(Vt) | 0, 20 mL ~ 1500 mL |
Frequence(Freq) | 1bpm ~ 100 bpm |
I/E | 2:1~1:6 |
Pressure Limit | 20 cmH2O ~ 100 cmH2O |
Monitored Parameters | |
Frequence(Freq) | 0 /min ~ 100 /min |
Tidal volume(Vt) | 0 mL ~ 2000 mL |
MV | 0 L/min ~ 100 L/min |
Graphical display: | |
P-T(pressure – time) | |
F-T(flow – time) | |
Size | |
1. Wooden case packing size:L 810*W 760*H 520mm , GW:53KG ; NW:30KG
2. Wooden case packing size:L 650*W 690*H 520mm , GW:40KG ; NW:20KG |
|
Alarm and protection | |
The AC power failure alarm | Power failure or no connection |
Internal battery backup low voltage alarm | <11.3±0.3V |
No tidal volume | ≤5Ml within 6s |
High oxygen concentration alarm
Low oxygen concentration alarm |
19%-100%
18%-99% |
High Airway pressure alarm
Low Airway pressure alarm High Minute Volume alarm Low Minute Volume alarm Continuous Pressure alarm |
20cmH2O-100cmH2O
0cmH2O-20cmH2O Adult(5L/min-20L/min) Paed(1L/min-15L/min 0-10L/min) (PEEP+1.5kPa) over 16s |
Suffocation warning | 5s-60s no spontaneous ventilation |
The maximum limited pressure | <12.5 kPa |
Fan error
Oxygen deficit |
Show on screen
Show on screen |
Working conditions | |
Gas source | O2 |
Pressure | 280kPa-600kPa |
Voltage | 100-240V |
Power frequency | 50/60Hz |
Input power | 60VA |